TROUBLESHOOTING
PF3000P
TROUBLESHOOTING
Problem
Possible Reasons
Corrective actions
LED is RED when switched
on
Low battery
Remove battery and
charge
Air flow too low
Remove HEPA filter and
replace
LED does not illuminate
when switched on
Battery is drained
Remove battery and
charge
Respirator is faulty
Return to service centre
for repair
Headband does not
tighten
Adjuster damaged
Replace headband
Ratchet damaged
Replace headband
Battery charger does not
switch on or charge the
battery
Mains electrical supply
not active
Check mains electrical
supply
Fuse in plug blown (if
applicable)
Replace fuse
Vehicle/site power supply
not active
Check vehicle/site power
supply
Fuse in 12V plug broken
Replace fuse
Battery damaged
Replace battery
Table 7
STORAGE (excluding battery)
The respirator should be stored in dry, clean conditions away from direct sunlight,
sources of heat, petrol and solvent vapours. Store in accordance with the guide below.
Respirator should be stored in the bag provided, when not in use.