Activate the *VNP polling mode, This mode is only available in ASCII mode
*VNP
Deactivate the *CV1 and *CV2 mode (Default mode at power up)
*RG1 /
*RG2
Return the current minimum and maximum range for channel 1 or 2.
The information is coded
The fist 5 byte is the identifier
00000: minimum scale.
00001: maximum scale.
10000: End of transmission.
The last 4 bytes are the scale coded in hex value, use the scale table bellow to
determine the scale
Example
After a RG1 you will receive
“:000000013
:00001001A
:100000000”
What it means
:000000013 : is the minimum scale 13 hex = 19 = 3m
:00001001A : is the maximum scale 1A hex=26=10 (Watt or Joule)
:100000000 : End of transmission.
*SO1 /
*SO2
Set Zero Offset of Channel 1 or 2.
This command subtracts the current value from all future measurements the moment
the command is issued to set a new zero point.
*CO1 /
*CO1
Clear Zero Offset of Channel 1 or 2.
This command undoes the Zero Offset command to set the zero point at zero.
*DO1 /
*DO2
Display Zero Offset of Channel 1 or 2.
This command displays the Zero offset value which.