Test Name
Standards
Test Specifications
Minimum
Performanc
e Criterion
Required
EUT
Serial
Number
Results
Surge Immunity
IEC61000-4-5 (2014)
Power: ±2kV L-PE / ±1kV L-
L
I/O Ports: ±1kV / N/A
Communication Ports: ±1kV /
N/A
B
Labcem#2340
Pass
Labcem#2341
N/A
Immunity to Conducted Disturbances,
Induced by Radio-Frequency Fields
IEC61000-4-6 (2013)
Power: 3V
I/O Ports: N/A
Communication Ports: N/A
A
Labcem#2340
Pass
Labcem#2341
N/A
Power Frequency Magnetic Field
Immunity
IEC61000-4-8 (2009)
Continuous Field: 30A/m /
50Hz & 60Hz
A
Labcem#2340
Pass
Labcem#2341
Pass
Voltage Dips, Short Interruptions and
Voltage Variation Immunity on AC
Input
IEC61000-4-11 (2004)
Voltage dips: 0%Un during 1
cycle
B
Labcem#2340
Pass
Voltage dips: 40%Un during
10 cycles (at 50 Hz)
B
Labcem#2340
Pass
Voltage dips: 40%Un during
12 cycles (at 60 Hz)
C
Labcem#2340
Pass
Voltage dips: 70%Un during
25 cycles (at 50 Hz)
C
Labcem#2340
Pass
Voltage dips: 70%Un during
30 cycles (at 60 Hz)
C
Labcem#2340
Pass
Short interruptions: 0%Un
during 250 cycles (at 50 Hz)
C
Labcem#2340
Pass
Short interruptions: 0%Un
during 300 cycles (at 60 Hz)
C
Labcem#2340
Pass
Table 8
Performance criteria
Performance
Criteria
Description
Criteria A
The DUT does not have any noticeable deviations in its performance before, during or after
the application of the EMC test.
If a range of performance is specified as normal operation, the DUT did not have any
readings outside of this range before, during or after the application of the EMC test.
Criteria B
The DUT does not have any deviations of performance before or after the application of the
EMC test, but during the application a deviation is noted that is not considered normal
operation. When the EMC test is paused or completed the DUT recovered on its own and
did not require any user intervention to return it to normal operation.
If a range of performance is specified as normal operation, the DUT’s performance was
outside of this range during the application of the test, but recovered on its own and operated
within its normal range after the application of the EMC test.
No permanent damage occurred to the DUT or any AE as a result of the disturbance.