background image

Features include broad temperature and high DS ratio. This
allows the user to monitor the target from a long distance;
far away from potential risk.

FEATURES:

• High DS ratio
• Adjustable emissivity from 0.1 to 1.00 in 0.01 steps
• Ultra low power consumption in shutdown mode
• Extended long time measuring reliability
• Laser sighting On/Off is switchable
• Backlit LCD display
• °F or °C selectable
• Electronic trigger lock
• Temperature data storage
• Audible alarms

APPLICATIONS:

• Manufacturing processes of semiconductor technology
• Automotive repair and maintenance
• Food safety and processing
• HVAC energy audits
• Electrical troubleshooting
• Testing terminals on circuits
• Scientific experiment
• Air conditioning

SAFETY INFORMATION:

Read the following safety information carefully before
attempting to operate or service the meter. Only qualified
personnel should perform repairs or services
not covered in this manual.

LASER WARNING NOTE!

Do not point laser directly at eyes.
Use caution around reflective surfaces.

Keep out of reach of children!

CAUTIONS!

DO NOT

submerge the unit in the water.

This product is not designed for use in medical evaluations.
The product can only be used to measure body temperature
for reference. It is meant for industrial and scientific
purposes.

SAFETY SYMBOLS

Dangerous, refer to this manual before using the
meter.
CE Certification

This instrument conforms to the following standards:
EN61326:

Electrical equipment for measurement, control and
laboratory use.

IEC61000-4-2:

Electrostatic discharge immunity test.

IEC61000-4-3:

Radiated, radio-frequency, Electromagnetic
field immunity test.

IEC61000-4-8:

Power frequency magnetic field immunity
test.

Tests were conducted using a frequency range of 80-
1000MHz with the instrument in three orientations.
The average error for the three orientations is ±1.0 °F
(±0.5 °C) at 3V/m throughout the spectrum. However,
between 781-1000MHz at 3V/m, the instrument may not
meet its stated accuracy.

The device may not be disposed in the trash. It
promotes the re-use recycling and other forms of

recovery of used materials and components, and to improve
the environmental performance of all operators
(manufacturers, traders, treatment facilities) involved in the
life cycle of products. Dispose of the product appropriately
in accordance with relevant regulations.

IRT675 Manual-FINAL2-050608:Layout 1  5/6/08  9:42 AM  Page 2

Summary of Contents for IRT675

Page 1: ...his is a non contact infrared temperature measuring instrument To measure a temperature point the unit at the desired object pull the measuring trigger and hold Make sure the target area is larger tha...

Page 2: ...product is not designed for use in medical evaluations The product can only be used to measure body temperature for reference It is meant for industrial and scientific purposes SAFETY SYMBOLS Dangero...

Page 3: ...Emissivity Adjustable 0 1 1 0 F C Switchable YES Backlight YES Laser Sight Switchable YES Audio Alarm YES Max Min Avg T YES Auto Measuring YES 10 Point Memory YES Dual Display YES Tripod Mount YES Di...

Page 4: ...e temperature within one shot measuring The highest and lowest temperature difference within one shot measuring Recall the stored data by using or button Emissivity for more accuracy on different mate...

Page 5: ...THERMOMETER FIELD OF VIEW FOV RATIO DISTANCE TO DIAMETER DS RATIO The field of view is the angle of vision at which the instrument operates and it is determined by the optics of the unit The FOV is th...

Page 6: ...d 127 260 0 06 Copper 199 390 0 18 599 1110 0 19 Steel 199 390 0 52 599 1110 0 57 Zinc galvanized sheet iron bright 28 82 0 23 Brass highly polished 247 476 0 03 Brass hard rolled polished w lines 21...

Page 7: ...es de temp rature Alarmes sonores APPLICATIONS Proc d s de fabrication de technologie de semi conducteurs R paration et entretien des automobiles S curit et traitement de l alimentation Audits nerg ti...

Page 8: ...ts Disposez donc de ce produit de mani re appropri e et conform ment la r glementation applicable SP CIFICATIONS Ratio distance point de mesurage 50 1 Gamme de temp rature 25 6 2372 F 32 1300 C Pr cis...

Page 9: ...ne 12 Valeur de r glage d enregistrement 13 Touche de verrouillage 14 Touche de mode 15 Bouton de mont e Mode de r glage des param tres Lecture et affichage de la temp rature durant le mesurage La tem...

Page 10: ...AROUGE RATIO DU CHAMP DE VISUALISATION FOV RATIO DE DISTANCE AU DIAM TRE DU POINT DE MESURAGE Le champ de visualisation est l angle de vision sous lequel l instrument fonctionne et il est d termin par...

Page 11: ...99 95 pur non oxyd 127 260 0 06 Cuivre 199 390 0 18 599 1110 0 19 Acier 199 390 0 52 599 1110 0 57 Fer en feuille recouvert 28 82 0 23 de zinc galvanis brillant Laiton hautement poli 247 476 0 03 Lai...

Page 12: ...C Seguro electr nico en el gatillo Memoria para dato de temperatura Alarmas sonoras APLICACIONES Procesos de fabricaci n con de tecnolog a de semiconductores Reparaciones y mantenimiento de autom vil...

Page 13: ...bricantes gremios y plantas de tratamiento involucrados en el ciclo de vida de los productos Deseche este instrumento apropiadamente de acuerdo a los reglamentos correspondientes ESPECIFICACIONES Rela...

Page 14: ...datos 4 T Emisividad 5 Alarma Temp Alta Baja 6 Tecla de memoria 7 Bot n para disminuir 8 SCAN HOLD AUTO MEDIR AUTO 9 Opci n F C 10 Pantalla principal de temperatura 11 M x M n Prom 12 Configurar Regi...

Page 15: ...debe usted estar de ste Cuando el di metro del objetivo es peque o es importante acercarle el term metro para asegurarse de medir s lo su temperatura y no la de sus alrededores EMISIVIDAD Es la capac...

Page 16: ...70 0 94 Hierro fundido derretido 1299 1399 3270 2550 0 29 Cobre pulido 21 117 70 242 0 02 Cobre brillante rascado no al espejo 22 72 0 07 Cobre placa fuertemente oxidada 25 77 0 78 Esmalte blanco fund...

Reviews: