background image

Fig. 1. The DT4947SD’s controls and indicators and other physical features

5

1-1 Liquid-crystal display 

1-2 POWER/ESC button 

1-3 HOLD/NEXT button

1-4 REC/ENTER button

1-5 Type 

button

1-6 T1-T2  

button

1-7 SET button

1-8 LOGGER OFFSET button

1-9 T1 through T4 thermocouple jacks

1-10 PT1 input jack

1-11 PT2 input jack

1-12 RS-232 output jack

1-13 Socket for 9VDC AC adapter

1-14 Tripod attachment nut

1-15 Battery compartment cover screws

1-16 Battery compartment cover

1-17 Kickstand

1-18 SD card socket

TOP

RIGHT

SIDE

BACK

BOTTOM

FRONT

Summary of Contents for DT4947SD

Page 1: ...DT4947SD USER S MANUAL 4 CHANNEL THERMOMETER WITH DATALOGGING SD CARD Please read this manual carefully and thoroughly before using this product...

Page 2: ...Setup 6 7 Adjusting Probe Offsets 8 Normal Operation 8 9 Holding and Storing Measurements 9 Automatic vs Manual Datalogging 9 10 Transferring Data from the SD Card to a Computer 10 11 Specifications 1...

Page 3: ...o 2192 F 100 to 1200 C using Type J thermocouples Type R and Type S thermocouples have an even higher upper limit These very wide ranges plus a fast response time and a low price tag suit thermocouple...

Page 4: ...tion Outputs Excel compatible data logs Accepts SD memory cards of up to 16 GB capacity Auto power off function Powered by six AA batteries or optional 9V AC DC adapter OPERATING INSTRUCTIONS WHAT S I...

Page 5: ...5 Type button 1 6 T1 T2 button 1 7 SET button 1 8 LOGGER OFFSET button 1 9 T1 through T4 thermocouple jacks 1 10 PT1 input jack 1 11 PT2 input jack 1 12 RS 232 output jack 1 13 Socket for 9VDC AC adap...

Page 6: ...l the word dAtE appears in the center of the display along with the flashing value 00 00 00 at the lower left above yy mm dd Set the current year by pressing the Type or T1 T2 button repeatedly until...

Page 7: ...ecrease or increase the value until the desired sampling time appears above SP t The options are 0 1 2 5 10 30 60 120 300 600 1800 and 3600 seconds 0 seconds to 1 hour 9 Format the SD card Once the sa...

Page 8: ...the LOGGER OFFSET button hold it for at least three seconds and then release the button A screen similar to the one at right should appear Next press the Type button which has the letter L stenciled o...

Page 9: ...you also can check the current date and time by briefly pressing the SET button which has the words Time check stenciled to its right Doing so causes both values to appear briefly at the lower left o...

Page 10: ...nd the term LOGGER to briefly appear at the upper right of the display As in automatic datalogging mode in this mode the instrument is storing measurements and their time stamps on the SD card In manu...

Page 11: ...11 Fig 2 Typical Excel data only screen Fig 3 Typical Excel graphics only screen...

Page 12: ...F or C above 1000 F for Types K J T and E 1 C or F for Types R and S Sampling time options 0 1 2 5 10 30 60 120 300 600 1800 or 3600 seconds 0 seconds to 1 hour Settable parameters Date time decimal...

Page 13: ...FORMATION General Tools Instruments General s DT4947SD 4 Channel Thermometer with Datalogging SD Card is warranted to the original purchaser to be free from defects in material and workmanship Subject...

Page 14: ...ntact our Customer Service to obtain an RGA Return Goods Authorization number before forwarding the unit via prepaid freight to the attention of our Service Center at this address General Tools Instru...

Page 15: ...__________________________________ __________________________________________________________________ __________________________________________________________________ _______________________________...

Page 16: ...31 6100 FAX 212 431 6499 TOLL FREE 800 697 8665 e mail sales generaltools com www generaltools com DT4947SD User s Manual Specifications subject to change without notice 2010 GENERAL TOOLS INSTRUMENTS...

Reviews: