Pressing the button twice displays the lowest reading
since
entering that measurement mode.
will appear at the left of
the primary readout.
•
Pressing the button a third time resumes real-time
measurement
and display of the same parameter.
RECALLING STORED READINGS
To enter Recall mode
, press the
<DN
and
UP>
buttons at the same time.
will appear
above the Record number indicator at the lower left of the LCD. The Record number indicator will
correspond to the last pair of readings stored.
You can now use the
<DN
and
UP>
buttons to navigate to and display a specific pair of readings.
For example, the screen shot below shows that a pair of readings of 63.8% for RH and 28.8°C for
temperature were stored in Record location 08.
Alternatively, you can repeatedly press the
UP>
button to recall all
readings in the order in which they were stored, or the
<DN
button
to recall them in reverse order. Pressing the
UP>
button when the
last stored pair of readings is displayed returns the Record
number indicator to
01
, corresponding to the first pair of readings
stored. Pressing the
<DN
button when the first stored pair of
readings is displayed returns the indicator to the record number of
the last pair of readings stored.
To exit Recall mod
e and resume making real-time measurements, press the
STORE/ENTER
button.
To erase all stored readings
, press and hold the
ZERO
and
UNIT
buttons at the same time for at least 2 seconds. The meter will
respond by sounding three short beeps and resetting the Record
number to
01
. You can clear all stored readings while operating in
Recall mode or Measurement mode. Clearing the memory in
Recall mode produces the display at right.
SETTING ALARMS
Note: Follow the instructions in this section to manually set, change or arm/disable the RH and
moisture level alarms using buttons on the meter. T
o perform the same tasks using the X-In-1
software
included with the meter, install and operate the program using the instructions in the
“Using the Included Software” section beginning on p. 15.
The RHMG700DL allows you to set alarms that will repeatedly
sound the beeper if the measured RH level or the moisture level of
a material (measured by either the pinless or pin-type sensor) is
above or below a certain value. The upper and lower limits are
called alarm setpoints. For example, the screen shot at right shows
that the upper limit for RH—the
HIGH
setpoint of the RH alarm—
has been set to 63.8%.
Wood moisture pro
Mini USB
L
L
9
Summary of Contents for RHMG700DL
Page 20: ...20...