
12
If the primary readout shows a value other than
0.0%
, you must
calibrate the meter to operate in
MI
mode. To do so, continue to
hold the meter in your hand (touching nothing) and press the
ZERO
button. Immediately after the term
ZERO
appears briefly below the
bar graph at the left of the display, the primary readout will show
0.0%
and the meter will be calibrated and ready to operate in
MI
mode.
To obtain accurate readings in
MI
mode, you must recalibrate the meter each time you re-enter
that mode from another mode.
CHECKING CALIBRATION
Although you cannot calibrate the RHMG700DL yourself, you can
check
the meter’s calibration
periodically to assure yourself of the accuracy of humidity-related readings (RH, absolute
humidity, dew point and mixing ratio). Factory calibration should minimize the humidity sensor’s
drift for several years—well beyond the warranty period.
At a minimum, a calibration check should entail immersing the slotted structure protecting the
humidity sensor in two bottles containing saturated salts. One reference salt should produce a
reading of a 33%; the other should produce a reading of 75%.
Two salts of this type are available from General: the HR33LJ (33%) MgCl Calibration Salt and the
HR75LJ (75%) NaCl Calibration Salt.
Depending on the size and shape of the calibration salt bottle that you use, you may have to
remove the hanger tab (Fig. 1, Callout B) from the meter to provide sufficient clearance for the
bottle to fit snugly over the slotted structure protecting the humidity/temperature sensor (after
removing the dust cap). The hanger tab is removable for that reason.
DATA LOGGING
This section explains how to use front-panel buttons on the RHMG700DL to view and manually
change data logging settings and to clear logged data from the meter's memory.
The final section of this manual—
Using the Included Software
, beginning on p. 15—outlines how
to do the same tasks more quickly and easily by pressing soft buttons on the X-In-1 application
included with the meter.
Viewing Data Logging Settings
To view your data logging settings
, press the
LOG/SET
button. If
the
▼
mode selector does not appear above
TM
, press the
<DN
or
UP>
button to move the selector above
TM
, at the left side of the
top line of the display. Once the selector is above
TM
, the two-digit
date portion of the system clock setting (09:18, or September 18,
2012 in the upper screen shot at right) will be visible on the
secondary readout.
Summary of Contents for RHMG700DL
Page 20: ...20...