CARE & MAINTENANCE
ü
Unplug the instrument from the power source – wall outlet.
ü
Wear appropriate safety protection.
ü
Clean inside hood when appropriate. A damp cloth is recommend for basic cleaning. Do not use solvents
on the instrument housing, sub-pan, keypad, power cord and power jack.
ü
Make sure no liquid enters the instrument housing.
ü
Handle with care to avoid damage to the weigh cell.
ü
Consult your safety officer with any questions or concerns.
ERROR MESSAGS & TROUBLESHOOTING GUIDE
Error codes appear on the display for 2 seconds. The program then returns automatically to the weighing mode
.
Display
Cause
Solution
No Segments on the
Display
No AC Power
The Power Supply is not
Plugged into the Balance
Blown Fuse
Check the AC Power Supply
Plug in the Power Supply
Replace Fuse
E _ _ _ _
The Weight Value Exceeds
the Capacity of the Balance
Remove Load from the
Weigh Pan
- E _ _ _ _
(
Under Load Indicator)
Improper Assembly of
Weigh Pan and Sub Pan
Assembly
Remove Weigh Pan, Sub
Pan and Reassemble
Err - - 0
&
Err - - 1
Calibration Failure
Environmental Factors
Causing Unstable
Conditions. Move to a More
Suitable Location
(see pages
2 and 5)
Inappropriate Calibration
Weight. Initiate Calibration
Procedure with Correct
Calibration Weight