Update will not proceed if the battery power is low. Please connect the
charger.
Do not turn off the power or remove the USB drive during an update.
Contact your manufacturer for the latest version of Firmware.
7.
Cleaning
Calibration and maintenance of equipment other than cleaning is only carried
out by the manufacturer. Use a soft cloth and 70% ethanol to remove
contaminants from the machine surface and tray except for LCD. In order to
clean LCD, please use a lens cleaner.
Make sure to turn off the power and use the liquid to prevent the solution
from infiltrating into the inside.
Validation of the device’s performance is possible by the user, but calibration
is only possible by the manufacturer and service center.
8.
Trouble shooting
Problem
Suggested Solution
In the event of no power 1. If the battery is completely discharged, charge for
about 30 minutes and try again.
2. If the circuit is broken by the unprovided adapters,
please contact the manufacturer for assistance.
** This machine does not support fast charging function.
In the event of frozen
screen while in use
1. Shut off the power and reboot using Forced Power off
switch on the side of the machine.
2. Please contact the manufacturer when the power
neither cannot be shut off nor reboot the machine.
In the event of RFID tags
reading failure.
1. Check if RFID tags attached to item boxes are broken.
2. Please contact the manufacturer if it recurs even after
reboot.
In the event of Invalid 1. Check if the test was carried out correctly according to