2.9 Uploading standard equation
This function shall be only used in case of occurring some fault in the download of
Internet curves of G939. In this function, the Standard Equations built into the G939 will
be uploaded, and they may not be the most updated. Moreover, there is no way to select
them, so all the equations will be uploaded.
The equations undergo updates with time, and we strongly recommend that a connection
Internet be made in order to download the most updated curves.
To upload Standard Equations:
1.
Connect the G939;
2.
Go to “Configuration” and depress YES; you have to put the Hard Lock in the serial
output;
3.
Go to “2.9 Uploading Standard Equation”, and depress YES;
4.
Use the Right hand/Left hand ARROWS to select the required language;
5.
Depress YES to complete.
How to exit “2.8 Uploading Internet Equation” function
This function passes the control of the G939 processor to the computer to which it
is connected. Once having entered in this function, we have only two alternatives
to exit:
A.
Exit with the site commands;
B.
Disconnecting, waiting for 5 seconds, and connecting again the power
supply to G939.