Programming Site Data
1-27
November 2004
Setting Up Output Scale
(cont.)
3.
Use the pull-down menu to select the desired channel option and
press
Next Item/Enter
.
4.
Use the pull-down menu to select the desired measurement
parameter and press
Next Item/Enter
. See Table 1-9 below for a
description of the available options.
5.
At
Zero
, enter a flow rate value for the low end of the analog
output range and press
Next Item/Enter
.
6.
At
Full
, enter a flow rate value for the high end of the analog
output range and press
Next Item/Enter
.
The
DIAG*
option listed above represents all of the individual
diagnostic parameters that appear at the previous prompt. See
Chapter 3,
Diagnostics
, of the
Service Manual
for a complete
description of these options.
Note:
The measurement units that appear in these prompts are those
selected in the
Global-System
window earlier in this section.
Procedure Options
After completing the above steps, IDM returns to the
Input/Output
window. Do one of the following:
•
To continue programming, refer to Appendix A,
Menu Maps
for IDM
, to navigate to the desired menu.
•
To leave the
User Program
, press
Exit Page
three times.
Table 1-9: Output Measurement Options
Options
Flow Velocity
Volumetric Flow
Forward Totalized Volume Flow
Reverse Totalized Volume Flow
Total Flow Measurement Time
Mass Flow
Forward Totalized Mass Flow
Reverse Totalized Mass Flow
Diagnostic Parameters*