Starting the System
4.4 Transducer
Connection
CAUTION!
Prior to connecting or disconnecting a probe freeze the image.
It is unnecessary to switch the unit off.
If a probe is disconnected while running (write mode) a software error may
occur. In this case switch the unit OFF and later ON (perform a reset).
CAUTION!
If the cable spout on the right-hand door is missing do not pull on the probe
cable, the probe cable can be damaged. Please insert the spout in the destined
place or call the Service Department.
1. Plug the probe connector into a free socket.
2. Twist the probe twist lock lever to vertical position to lock
it in place.
3. Open the right-hand side door.
4. Lay the cable into the intended cable holder and close the
door.
Each probe connector has a mechanical fixation lock which has to be engaged to enable the probe to
be working at all.
Probe Connector:
shown unlocked. Turn the knob clockwise to lock.
Note:
A disconnection of an active probe is permitted in read mode
only !!!
If a probe is disconnected while running (write mode) a software error can occur!
Probe holders:
Always store the probes in the intended probe holders.