Starting the System
4
Starting the System
4.1 General
Remarks
Installation, switching on for the first time and check-up of the system must only be performed by
authorized persons.
The VOLUSON
®
730 is delivered with recommended basic settings. These offer suitable conditions
for a large number of applications. Depending on the user's experience these default settings can be
changed and stored as new User Programs. Storing these programs or quick loading of new programs
of a second user is done by back-ups.
4.2 Safety
Warnings
CAUTION ! In the event the equipment has been brought from cold environment (stock
room, airfreight) into a warm room, allow several hours for temperature balance
and passing of condensation humidity before switching on the first time (risk of
leakage current).
CAUTION ! The system is equipped with mains outlets separated by an isolation transformer
for peripheral equipment (printer, VCR). To ensure electrical safety, these
instruments must never be connected to a wall socket.