Rendered PDF File Page 36 of 73
DOC1710900, Rev:1
Query Retrieve
Configure
Configure – Query Retrieve
From dataflow name
pulldown list, select the
DICOM
Query retrieve
dataflow
Select
DicomQR
from
the Dataflow Overview
window
Select
Properties
Required
fields
The
Query Retrieve
Properties window will appear. The
required fields must be filled out. Other fields might
need to be adjusted
The DICOM Query Retrieve is an
Input
device in the
Workflow/Dataflow, and the Vivid must be registered as a user on the
Q/R DICOM Server. Dicom QR configuration on EchoPAC might also
require Firewall modifications
Enter correct
AE title,
as it is
case sensitive and
Port No
.
IP Addr
must be selected
from the pulldown menu
and
Name
is optional, but
recommended
If
IP Addr
is not in pulldown
menu, go to
TCPIP tab
and
select
Modify
to change or
Add
.
A
Server Config
window will
appear
Insert a
descriptive
Name
and
IP
Address
Check
will ping the server
Select
Add
to
insert a new,
select
Modify
to
correct
IP Addr
of
existing DICOM
Device
DOC1710900 rev 01
Released