Section 3-2 - Installation Reminders
Section 3-2
Installation Reminders
3-2-1
Average Installation Time
The Vivid q N installation and functional checkout will take approximately one hour; Vivid q Nconsoles
with optional equipment may take slightly longer.
Once the site has been prepared, the average installation time required is shown in
below.
3-2-2
Installation Warnings
1.) The Vivid q N portable ultrasound scanner weighs only 5 Kgs (11 lbs). (This is the weight of the
scanner without any optional accessories). However since the SafeLock Cart weighs approximately
40 kg (88 lbs), two persons are always required to unpack it.
2.) There are no operator-serviceable components. To prevent shock, do not remove any covers or
panels. Should problems or malfunctions occur, unplug the power cord. Only qualified service
personnel should carry out servicing and troubleshooting.
3-2-2-1
System Acclimation Time
Following transport, the Vivid q N system may be very cold, or hot. Allow time for the system to
acclimate before being switched ON. Acclimation requires 1 hour for each 2.5
o
C increment, when the
temperature of the system is below 10
o
C or above 35
o
C.
Table 3-2
Average Installation Time
Description
Average
Installation Time
Comments
Unpacking the scanner
20 minutes
Installing the scanner
30 minutes
Time may vary, according to the required configuration
DICOM Option
(connectivity)
30 minutes
Time may vary, according to the required configuration
CAUTION
Turning the system ON after arrival at the site - without allowing time for acclimation - may cause
system damage!
Table 3-3
Vivid q N System Acclimation Time
ALLOWED TRANSPORTATION AND STORAGE TEMPERATURES
OUT OF SPEC!
°
C
60
55
50
45
40
35
30
25
20
15
10
5
0
-5
-10
-15
-20
-25
-30
-35
-40
°
F
140
131
122
113
104
96
86
77
68
59
50
41
32
23
14
5
-4
-13
-22
-31
-40
Hrs
8
6
4
2
0
0
0
0
0
0
0
2
4
6
8
10
12
14
16
18
20