Section 2-2 - General console requirements
2-2-2-6
Power stability requirements
Voltage drop-out
Max 10 ms.
Power transients
(all applications)
Less than 25% of nominal peak voltage for less than 1 millisecond for any type of transient, including
line frequency, synchronous, asynchronous, or aperiodic transients.