Data Output
24
TONOPORT VI
2001589-312 Revision B
5
Data Output
The measurement data are output via CASE/CardioSoft.
Put the PC-based system into operation
(see Operator Manual of CASE, CardioSoft).
Turn off TONOPORT VI.
Connect TONOPORT VI to the PC system:
– via cable 2001589-040 if the USB port of
TONOPORT VI is used (
b
– via cable 2001589-011 if the serial port of
TONOPORT VI is used (
a
Turn on TONOPORT VI and wait for the time to be
displayed.
For more information about data output, please refer to
the Operator Manual of CASE, CardioSoft.
When you have finished downloading data to CASE/
CardioSoft and do not intend to continue working with
this system, disconnect TONOPORT VI and turn it off.
Fig. 5-1 Connections for PC cable
a
RS232 port
b
USB port
Warning
Risk to Persons—
Disconnect TONOPORT VI from other equipment
(CASE, PC) before connecting it to the patient.
Note
If the USB port is used (CardioSoft only), it is
necessary to first install the appropriate driver (see
"Software Installation").
CASE
must always be connected to the serial port.
b
a