Electromagnetic Compatibility (EMC)
34
TONOPORT VI
2001589-312 Revision B
Recommended separation distances between portable and mobile RF communications equipment and
TONOPORT VI
TONOPORT VI is intended for use in the electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of TONOPORT VI can help prevent electromagnetic interference by maintaining
a minimum distance between portable and mobile RF communications equipment (transmitters) and TONOPORT VI
as recommended below, according to the maximum output power of the communications equipment.
Rated maximum output
power of transmitter
[W]
Separation distance according to frequency of transmitter
[m]
150 kHz to 80 MHz
d = 1.17
80 MHz to 800 MHz
d = 1.17
800 MHz to 2.7 GHz
d = 2.33
0.01
0.12
0.12
0.24
0.1
0.37
0.37
0.74
1
1.17
1.17
2.34
10
3.69
3.69
7.38
100
11.67
11.67
23.34
For transmitters rated at a maximum output power not listed above, the recommended separation distance
d
in meters
(m) can be estimated using the equation applicable to the frequency of the transmitter, where
P
is the maximum
output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all instances. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.