Setup
2001589-312 Revision B
TONOPORT VI
17
Before using TONOPORT VI on a patient
1.
clear the memory
2.
check date and time and adjust if required
3.
select a measurement protocol
4.
enable or disable the audio signal.
Clearing the Memory
The symbol
M
on the display indicates that the memory
holds BP data. If these data still need to be analyzed, refer
to chapter 5 "Data Output" for details on data evaluation.
If you do not need the data any more, delete it as follows:
Briefly switch TONOPORT VI off and on again and
wait for the time to be displayed.
Push
: the display indicates "H 1".
Push
: the display indicates "LLLL".
To delete the data, push
again: the display
indicates "0000", followed by the time (if you do not
wish to clear the memory, turn off the BP monitor
instead of pushing
).
Selecting the Measurement Method
Briefly switch TONOPORT VI off and on again and
wait for the time to be displayed.
Push
9 times: the display indicates "H 9".
Push
: the display indicates "0000" if the
selected method is the deflation measurement
method, or "1111" if the selected method is the
inflation measurement method.
Either confirm with
or switch to the other op-
tion with
, then confirm with
.
Note
When using TONOPORT VI in conjunction with
CASE/CardioSoft, it is recommended to perform
the first three steps at the PC.