Ultrasound
Pulser type
Spike pulse
Pulse repetition frequency (PRF) max. 1,000 Hz, 3 settings (depending on the calibration range),
continuous setting approx. 8 ... 1,000 Hz., no continuous pulse sequence
Voltage
max. 400 V
Charging capacitor
1 nF, 220 pF
Initial pulse rise time
< 15 ns
Damping
50 Ohm, 500 Ohm, 1 kOhm
Probe mode (TR)
On/Off, through-transmission
Settings and evaluation
Calibration ranges
min. 0 ... 2.5 mm; 0 ... 0.1" (steel); max. 0 ... 9,700 mm; 0 ... 381" (steel)
Sound velocity range
500 ... 15,000 m/s; 0.02 ... 0.59 "/
μ
s; integrated, editable material table
Pulse shift
–10 ... 1,500 mm; –0.39 ... 50" (steel)
Probe delay
0 ... 100
μ
s
Frequency range
0.5 ... 20 MHz (–3 dB); 4
fi
lter ranges
Gain
110 dB, adjustable in steps of 0.5/1/2/6 dB
Specifi cations SpotChecker
Technical
speci
fi
cations
SpotChecker
Issue 02, 03/2009
79
Summary of Contents for SpotChecker
Page 1: ...SpotChecker Technical Reference and Operating Manual Ident No 49 131 Rev 2 0 ...
Page 2: ...This Issue 02 03 2009 applies to the version 1 0 Subject to change without notice ...
Page 7: ...Introduction 1 SpotChecker Issue 02 03 2009 7 ...
Page 15: ...Important information on ultrasonic testing Introduction SpotChecker Issue 02 03 2009 15 ...
Page 17: ...The SpotChecker Introduction SpotChecker Issue 02 03 2009 17 ...
Page 24: ...24 Issue 02 03 2009 SpotChecker ...
Page 25: ...Initial start up and basics 2 SpotChecker Issue 02 03 2009 25 ...
Page 48: ...48 Issue 02 03 2009 SpotChecker ...
Page 49: ...Individual inspection reports 3 SpotChecker Issue 02 03 2009 49 ...
Page 70: ...70 Issue 02 03 2009 SpotChecker ...
Page 71: ...Care maintenance errors 4 SpotChecker Issue 02 03 2009 71 ...
Page 74: ...74 Issue 02 03 2009 SpotChecker ...
Page 75: ...Technical specifications 5 SpotChecker Issue 02 03 2009 75 ...
Page 82: ...82 Issue 02 03 2009 SpotChecker ...
Page 83: ...Annex 6 SpotChecker Issue 02 03 2009 83 ...
Page 92: ...Annex Disposal 92 Issue 02 03 2009 SpotChecker ...