Racer view
Racer view with test sketch
Automatic A-scan freeze
SpotChecker can automatically freeze the A-scan by
means of two-dimensional gates. Up to three gates
can be set up for this purpose. As soon as the echoes
of the currently active A-scan are situated within these
gates, the A-scan is automatically frozen. The posi-
tion of the
fi
rst gate must ensure that the
fi
rst backwall
echo reaches 80 % screen height.
H
Note:
More information on the positioning of gates are given
in the on-line help supplied with SpotChecker.
The software SpotChecker
Individual inspection reports
SpotChecker
Issue 02, 03/2009
65
Summary of Contents for SpotChecker
Page 1: ...SpotChecker Technical Reference and Operating Manual Ident No 49 131 Rev 2 0 ...
Page 2: ...This Issue 02 03 2009 applies to the version 1 0 Subject to change without notice ...
Page 7: ...Introduction 1 SpotChecker Issue 02 03 2009 7 ...
Page 15: ...Important information on ultrasonic testing Introduction SpotChecker Issue 02 03 2009 15 ...
Page 17: ...The SpotChecker Introduction SpotChecker Issue 02 03 2009 17 ...
Page 24: ...24 Issue 02 03 2009 SpotChecker ...
Page 25: ...Initial start up and basics 2 SpotChecker Issue 02 03 2009 25 ...
Page 48: ...48 Issue 02 03 2009 SpotChecker ...
Page 49: ...Individual inspection reports 3 SpotChecker Issue 02 03 2009 49 ...
Page 70: ...70 Issue 02 03 2009 SpotChecker ...
Page 71: ...Care maintenance errors 4 SpotChecker Issue 02 03 2009 71 ...
Page 74: ...74 Issue 02 03 2009 SpotChecker ...
Page 75: ...Technical specifications 5 SpotChecker Issue 02 03 2009 75 ...
Page 82: ...82 Issue 02 03 2009 SpotChecker ...
Page 83: ...Annex 6 SpotChecker Issue 02 03 2009 83 ...
Page 92: ...Annex Disposal 92 Issue 02 03 2009 SpotChecker ...