Chapter 4: Installation
Table 3: Analog Outputs (TB1)
Pin
Signal
Function
1
Analog 1 (+)
4-20mA Stream #1
2
GND_ISO (-)
3
ANALOG 2 (+)
4-20mA Stream #2
4
GND_ISO (-)
5
ANALOG 3 (+)
4-20mA Stream #3
6
GND_ISO (-)
7
ANALOG 4 (+)
4-20mA Stream #4
8
GND_ISO (-)
9
ANALOG 5 (+)
4-20mA Stream #5
10
GND_ISO (-)
11
ANALOG 6 (+)
Stream ID
(11 mA, 12 mA, 13 mA, 14 mA, or 15 mA, as
12
GND_ISO (-)
N
OTE
:
4-20 mA outputs are isolated from system ground.
4-20 mA outputs are not individually isolated from each other and, therefore, are not intended to be
connected to multiple systems with differences in ground potential.
4-20 mA outputs may be configured by the user to output NPOC, IC, TC, or TOC ppm levels.
Table 4: Stream ID
Active Stream
Stream ID
Standby
Output matches the
standby
level set for the first
sequentially-ordered active stream, starting with Stream 1
1
11 mA
2
12 mA
3
13 mA
4
14 mA
5
15 mA
Error
Output matches the
error
level set for the first
sequentially-ordered active stream, starting with Stream 1
Warning
n/a (Does not change current value.)