Chapter 6: Analyzer Operation
Continuous Measurements
By turning
Continuous
On
, the Analyzer makes continuous measurements of the enabled sample streams and
the Check Standard (if it is enabled). Continuous measurement of Stream 1 is shown in
. Only Stream 1
and the Check Standard are enabled in this example, so measurements of Stream 1 are made without a pause,
with two exceptions.
Figure 41: Continuous Operation
The first exception is the Check Standard, since it is enabled. Measurements of the Check Standard are made
according to the schedule established by the operator. In the example shown in
, the check standard is
measured every three hours (the Check Standard Frequency). Since two
Repetitions
have been selected, the
check standard is measured twice every three hours. The first measurement is ignored because it has been
designated as a
Reject
.
Figure 42: Operation with a Cycle Frequency of One Hour