14
RPT 301 User Manual
K177 Issue No. 3
2.
U, O; G; R; G; R; U, 16; G; R; G; R; G; R; A, 100
Reply: Reading 1 mbar
Reading 2 mbar
Reading 3 psi
Reading 4 psi
Reading 5 psi
Then at 100 second intervals - Reading psi
This command string sets the instrument units to mbar, returns two
consecutive readings, changes units to psi, returns three consecutive
readings and then selects auto-send mode with a period of 100 seconds
between readings.
Command String Limitations
Auto-send mode and diagnostics mode are cancelled when a further
command is received, this should be used at the end of a command
string.
Note:
The C, L and T commands are interactive and are not
suitable for multiple commands.