24
RPT 301 User Manual
K177 Issue No. 3
The calibration routine requires a minimum of two pressures within the
operating extremes (maximum of six pressures). The instrument
software calculates an offset and gain correction to be applied to all
subsequent readings using a statistical best straight line method.
For single point calibration, the gain factor is set to unity.
Notes
1
For the pressure sensor to rapidly respond to the applied pressure,
disable the filter by sending the filter command F, 0,0<CR>.
2
Before sending the ‘C’ command set the instrument to operate in the
pressure measurement units required for calibration. i.e. U, 0<CR>
selects mbar. Always use the same units when entering applied
pressures.
Command:
C, PIN
Reply:
APPLY PRESSURE nn
apply pressure, when steady send <CT>
Reply:
ENTER APPLIED PRESSURE nn
type pressure value - decimal number followed by
<CT>
Reply:
MORE? (Y/N)
type Y <CT> to calibrate more points
or
type N <CT> to complete calibration point entry
Reply (
if N
):
SLOPE XX INTERSECT YY
ACCEPT CALIBRATION (Y/N)
type Y
<CT>
to accept and permanently update the calibration
or
type N <CT> to terminate without saving
Reply
(if N):
CALIBRATION TERMINATED
end of calibration procedure
Reply (if Y):
ENTER CAL DATE
type date in the format DD/MM/YY
Reply:
CALIBRATION COMPLETE
Where:nn = pressure point 1-6
Summary of Contents for Sensing RPT 301
Page 1: ...RPT 301 Digital Output Pressure Transducer User Manual K177 GE Sensing ...
Page 2: ... General Electric Company All rights reserved ...
Page 4: ...K177 Issue No 3 ...
Page 8: ...K177 Issue No 3 iv RPT 301 User Manual ...
Page 13: ...K177 Issue No 3 RPT 301 User Manual 5 Figure 2 2 Instrument Dimensions ...