GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Job Card D/R A312 - Mouse
Page no. 1417
Chapter 9
JC-DR-A-312.fm
Job Card D/R A312 - Mouse
Chapter 9
1
SUPPLIES
•
Keyboard type 6 UNIX and Mouse SUN
for V2 Control Stations,
Keyboard type 7 UNIX and Mouse
for V3 Control Stations,
for V4 Control Stations
•
Black cable ties
2
APPLICABILITY
Three types of Rotative Arm Sets exist, as follows:
•
Ergotron A (introduced with V1 Control Stations) - contains one rotative arm support and two exten-
sion (rotative) arms
•
Ergotron B (introduced with V2 Control Stations and early V3 Control Stations) - contains one rotative
arm support and two extension (rotative) arms
•
Sisfle (introduced with V3/V4 Control Stations) - contains one rotative arm support and one extension
(rotative) arm
The procedure used in this Job Card vary according to the type of rotative arm you have.
Since the mouse is not supported on Ergotron A arms, this Job Card is only applicable to V2/V3/V4 Con-
trol Stations.
3
TOOLS
4
REQUIRED EFFORT
Personnel: 1 Field Engineer
Time: 30 minutes
5
SAFETY PRECAUTIONS
WARNING
Before beginning work, shut down correctly the equipment and switch off the Senographe
system hospital supply circuit breaker. Lock it in the off position or mark it clearly with a
warning sign to prevent others from applying power
.
6
PREREQUISITES
None