MAINTENANCE
GE Medical Systems
SENOGRAPHE 800T
REV 6
om 2135575–100
3–20
4–5
Minimum Optical Density Selection
A target Optical Density has been programmed, based on your recommendations, by the GE
field service engineer during the installation of your Senographe 800T. The machine will
never allow an exposure if this target optical density cannot be reached, except if a maximum
allowable drop in optical density has been programmed. If this is the case, then, in case the
target optical density cannot be reached, the machine will allow the optical density to drop by
the programmed value.
Example: If the target optical density has been set at 1.6, and a maximum allowable drop of 0.2
has been programmed, then the Senographe 800T will allow the optical density to drop down
to 1.6 – 0.2 = 1.4 if the target optical density cannot be reached.
In order to guarantee best results, the programming of this maximum allowable drop must
done by the GE field service engineer. If you do not wish to allow any drop in optical density,
then a value of zero must be programmed for the maximum allowable drop.
4–5–1
Access to Programming Mode
To access the programming mode, press
.
The APPLICATION Menu, focal spot, focal track, filter, etc., disappears.
4–5–2
Selecting a Function
To select a function, press the key located under the symbol for that function.
Summary of Contents for SENOGRAPHE 800T
Page 6: ...GE Medical Systems SENOGRAPHE 800T REV 6 om 2135575 100 iv Blank page ...
Page 12: ...GE Medical Systems SENOGRAPHE 800T REV 6 om 2135575 100 x Blank page ...
Page 14: ...GE Medical Systems SENOGRAPHE 800T REV 6 om 2135575 100 xii Blank page ...
Page 16: ...GE Medical Systems SENOGRAPHE 800T REV 6 om 2135575 100 xiv Blank page ...
Page 33: ...INTRODUCTION GE Medical Systems SENOGRAPHE 800T REV 6 om 2135575 100 1 17 Blank page ...
Page 38: ...INTRODUCTION GE Medical Systems SENOGRAPHE 800T REV 6 om 2135575 100 1 22 Blank page ...
Page 80: ...SPECIFICATIONS GE Medical Systems SENOGRAPHE 800T REV 6 om 2135575 100 4 6 Blank page ...
Page 82: ...REGULATIONS GE Medical Systems SENOGRAPHE 800T REV 6 om 2135575 100 5 2 Blank page ...