g
Events, descriptions, and states are shown in Table 6. Each event has one or more possible
states.
In the cases where more than one state is possible, an update to the event log and/or a
notification will be sent if the state changes, assuming the event were configured to do so.
For those events with only one state such as Date/Time Set, the log or notification is
written or activated whenever this action occurs, even though there is no change in state.
Table 6 Events
Event
Event Description
State
Range
Indicates the current
electrometer range.
RangeLow
RangeHigh
ElectrometerUpdate
This event is set
whenever a new dose
rate measurement is
available.
ElectrometerUpdate
Electrometer EepromStatus
At startup and upon
command an EEPROM
test can be run and the
results are reported
with this event.
EEPROMIdle
EEPROMPass
EEPROMFail
EEPROMInvalidConfiguration
Electrometer ConfigurationChange Set if any change to the
electrometer
configuration occurs.
Only used for event logging and
will never change on this screen.
HPICConfigurationChange
Set if any change to the
HPIC configuration
occurs.
Only used for event logging and
will never change on this screen.
External Power
Detects the presence or
absence of external DC
power.
Power Off
Power On
BatteryPresent
Indicates whether a
battery is installed or
not.
BatteryPresent
BatteryNotPresent
SystemStatus
A global indication as
to whether alarms are
active.
SystemStatusNormal
SystemAlarm
DatabaseStatus
Current database
status.
DatabaseOK
DatabaseError