page 37
© 2003 General Electric
STEP 4: POST RESCUE
After transferring the patient to ALS personnel, prepare the AED for the next rescue:
1.
Retrieve the rescue data stored in the internal memory of the AED by using
RescueLink software installed on a PC (see detailed procedure in the Data
Management section).
2.
Erase the internal memory of the AED.
3.
Connect a new pair of electrodes to the AED.
4. Close the lid.
5.
Verify that the
STATUS INDICATOR
on the Responder AED handle is
GREEN
.
WARNINGS
The following cautions must be observed to prevent problems during the rescue.
DANGER
: Fire and Explosion Hazard
Exercise caution when operating the AED close to flammable gases (including
concentrated oxygen) to avoid possible explosion or fire hazard.
WARNING:
Shock Hazard
Defibrillation shock current flowing through unwanted pathways is potentially a serious electrical
shock hazard. To avoid this hazard during defibrillation abide by all of the following:
•
Do not touch the patient, unless performance of CPR is indicated
•
Do not touch metal objects in contact with the patient
•
Keep defibrillation electrodes clear of other electrodes or metal parts in contact with patient
•
Disconnect all non-defibrillator proof equipment from the patient before defibrillation
WARNING:
Shock and Possible Equipment Damage
Disconnect all non-defibrillator proof equipment from the patient before defibrillation to prevent
electrical shock and potential damage to the equipment.
WARNING:
Electric Shock and Fire Hazard
Do not connect any telephones or unauthorized connectors to the socket on this equipment.
CAUTION:
Use only Approved Equipment
Using batteries, electrodes, cables, or optional equipment other than those approved by
General Electric may cause the Responder AED to function improperly during a rescue.
CAUTION:
Possible Improper AED Performance
Using electrodes that are damaged or expired may result in improper AED performance.
Section 4 - Instructions for Use
201-9198-001 Rev A
Summary of Contents for Responder AED
Page 1: ...Operation and Service Manual ...
Page 15: ...page 14 2003 General Electric 201 9198 001 Rev A ...
Page 41: ...page 40 2003 General Electric 201 9198 001 Rev A ...
Page 58: ......