SCHEDULED MAINTENANCE
DAILY MAINTENANCE
Check the
STATUS INDICATOR
to ensure that it is
GREEN
. When the indicator is
GREEN
, the
Responder AED is ready for a rescue. If the indicator is
RED
, refer to the Troubleshooting Table
in this chapter.
MONTHLY MAINTENANCE
1.
Open the AED lid.
2.
Wait for the AED to indicate status:
Observe the change of the
STATUS INDICATOR
to
RED
. After less than 5 seconds, verify that the
STATUS INDICATOR
returns to
GREEN
.
3.
Observe the expiration date on the electrodes.
4.
Listen for the voice prompts.
5.
Close the lid and confirm that
STATUS INDICATOR
remains
GREEN
.
ANNUAL MAINTENANCE
Perform the following tests annually to confirm that the diagnostics are functioning properly and to verify the
integrity of the case.
Check the Integrity of the Electrodes and Circuitry
1.
Open the AED lid.
2.
Remove the electrodes.
3.
Close the lid.
4.
Confirm that the
STATUS INDICATOR
turns red.
5.
Open the lid and confirm that the
ELECTRODE
indicator is lit.
6.
Reconnect the electrodes and close the lid.
7.
Make sure the expiration date is visible through the clear window of the lid.
Check to make sure that the
STATUS INDICATOR
is
GREEN.
8.
Open the lid and confirm that no diagnostic indicators are lit.
9.
Check the expiration date of the electrodes; if expired, replace them.
10. Check the electrode’s packaging integrity.
11. Close the lid.
Section 6 - Maintenance & Troubleshooting
201-9198-001 Rev A