STATUS LEDS
There are three system status LEDs on the front panel of the Responder 2000
AC Power (Green LED)
The AC Power LED is lit when the Responder 2000 is connected
to external AC power.
Battery Charging (Yellow LED)
The Battery Charging LED is lit when the Responder 2000
Battery is charging in the Responder 2000 or the battery charge
is being maintained.
Service Required (Red LED)
The Service Required LED is lit when the Responder 2000
requires service. Please take the Responder 2000 out of service
and Contact Customer Service.
PADDLE CONTROLS
The Apex paddle has one button, which controls both charge and shock. Press the Apex paddle button to charge the
Responder 2000. After the Responder 2000 is charged, press both the Apex and Sternum paddle buttons simultaneously
to deliver the shock. Charge and Shock are only activated when the system is in the proper defibrillation mode. If a
button is stuck or remains pressed from before activation of defibrillation, it must be released before a further press is
accepted. When the paddles are connected to the Responder 2000, the charge and shock buttons on the front panel will
be disabled. The charge button is enabled only when the Responder 2000 is in Manual Mode.
RS-232 DATA TRANSFER CONNECTION
This feature is used by factory authorized personnel only.
Summary of Contents for Responder 2000
Page 1: ...Operator s Manual 2026116 001 Revision B English ...
Page 8: ...2026116 001 Revision B Responder 2000 Page 8 ...
Page 46: ...2026116 001 Revision B Responder 2000 Page 46 ...
Page 118: ...2026116 001 Revision B Responder 2000 Page 118 ...
Page 141: ...2026116 001 Revision B Responder 2000 Page 141 ...
Page 142: ...Section 7 Accessories 2026116 001 Revision B Responder 2000 Page 142 ...