Programming Site Data
3-15
August 2003
Entering Probe Identification Data (cont.)
6.
The final prompt asks you to select a time period at which the
screen displays a calibration reminder for the probe,
determined from the last calibration date. GE Panametrics
recommends recalibrating a probe annually.
a.
Press
[ENTER]
to open the drop-down list. The menu
offers a choice of five time periods:
•
None
•
6 months
•
12 Months
•
18 Months
•
24 Months
b.
Scroll to the desired period, and press
[ENTER]
to confirm
the entry.
7.
Press the [
S
] key to return to the Probe ID tab.
You now have three options:
•
Use the arrow keys to enter another Calibration window and
program additional data.
•
To confirm the data and return to Operate mode, press
[F3]
(OK). The PM880 returns to Operate Mode.
•
To leave the window without confirming the data, press
[F2]
(Cancel) or the
[ESC]
key. The PM880 returns to Operate
Mode.