Appendix D. Foundation Fieldbus Communication
72
PanaFlow™ XMT1000 User’s Manual
D.6.1 Clearing the Totalizer
Batch totals can be controlled through Foundation Fieldbus (see
below). The user can start, stop, or reset
batch totalizers by setting the option on the
TOTALIZER_CMD
function of the
Composite Transducer Block
. To set the
totalizers from the Foundation Fieldbus:
1.
Verify that the gate and ground terminals on the Main PCB are connected.
2.
Program the
CALIBR_MODE_SELECTION
parameter on the
Composite Transducer Block
to
Gate Input
.
After these steps are complete, you can control the batch totalizer (start, stop, or reset) by selecting the desired option
on
TOTALIZER_CMD
and writing the changes to the meter. No password is required for this function.
IMPORTANT:
The
Inventory Totalizer
can only be reset at the factory.
Figure 29: TOTALIZER_CMD Function on Composite TB