background image

1.

  

Thank you.

  Thank you for purchasing the 

PanaFlow HT ultrasonic flowmeter.  Before removing 

the PanaFlow HT system from the crate, please inspect 

the flowmeter.  Each instrument manufactured by GE 

Measurement & Control is warranted to be free from 

defects in material and workmanship.  Before discarding 

any of the packing materials, account for all components 

and documentation listed on the packing slip.  

2.

  

Meter Installation

.  The ideal site is a straight section 

of process pipe that meets the following criteria:

•  Pipe section is horizontal and above ground.

•  Pipe section can accommodate the overall length of the 

flowcell.

•  There is sufficient clearance for the transducers on 

the sides of the meter (at the 3 o’clock and 9 o’clock 

positions).  

•  There are at least 10 pipe diameters of straight, 

undisturbed flow upstream and 5 pipe diameters 

of straight, undisturbed flow downstream from the 

transducers.(See 

User’s Manual

.)

•  For remote-mounted electronics, the electronics 

console should be located within 100 feet (30 

meters) from the meter body.  (See 

User’s Manual

.)

5

Communication and Power Wiring.

  The picture below 

shows the communication and power wiring to XMT900 

electronics.  Make all communication wiring before wiring 

the power terminals.

 

Once all wiring is complete, apply the power.

6.

 

 Programming & Testing.

  

The keypad and magnetic wand enable 

programming of the instrument through the glass 

faceplate without removing the cover. Thus, all 

programming procedures may be performed while 

the unit is installed in a hazardous area.  See the 

User’s Manual

 for instructions on programming.

PIN   Label

                             

Description

1

                                             

A-

                                               

Analog Output A:4-20mA/HART Output(SIL) Negative     

2

                                             

A+

                                               

Analog Output A:4-20mA/HART Output(SIL) Positive

3

                                             

B-

 

                                         

Analog Output B:4-20mA Negative

4

                                             

B+

                                              

Analog Output B:4-20mA Positive

5

                                             

C-

                                              

Digital Output C Negative

6

                                             

C+

                                              

Digital Output C Positive 

7

                                              

D-

 

                                          

Digital Output D Negative

8

                                              

D+

                                             

Digital Output D Positive

Terminal Block - Power Connection(AC)

PIN  Label

                             

Description                               

1

                                   

L2/N

                              

Neutral Connection

2

                                   

L1

                                                   

Live Power Connection

Terminal Block -Modbus/Calibration Connections

PIN  Label

                                 

Description

1

                                  

MOD-

                            

Modbus/Service Port Negative

2

                                  

MOD+

                            

Modbus/Service Port Positive

3

                                  

CAL-

                                 

Calibration Port Negative

4

                                  

CAL+

                                 

Calibration Port Positive

WARNING!   The PanaFlow HT flow transmitter can measure the flow rate of many fluids, some 
potentially hazardous. Be sure to follow local safety codes and regulations for installing electrical  
equipment and working with hazardous fluids or flow conditions.