Illustration 2-40 : Standard Table & Manual WS at Foot with 3M bridge
Item
Limit
Dimension
Recommend
Dimension
Description
mm
inch
mm
inch
(1)
Min
630
24.8
700
27.6 Foot end of rail to center line of WS detector.
With manual WS, the Min value should be larger
than 630 mm (530+100 Cal space)
(2)
Min
203
8.0 400
15.7
Service
clearance for WS installation.
The Min value should be larger than 203 mm, the
WS back cover could be installed or can be
removed.
(3)
Min
2060
81.1
2300
90.6 Back of bridge to center line of WS.
1.With manual WS, the Min value should be
larger than 2060 (860 +100 Cal space + 700
distance between focal spot to back of OTS
400 half table Width)
2.with manual WS, the Max value should be less
than 2735 ( OTS biggest travel range2835 - 100
Cal space)
Max
2735
107.7