Routine Maintenance
4.5.1 Product Water Culture Procedure
This procedure assumes the RO is connected to power and water, and is in operating
condition.
1.
Connect the product rinse line to the product port. Place the other end in a drain.
2.
Connect the waste line to the waste port. Place the other end in a drain.
3.
Turn on the RO unit. Allow the RO unit to run for 10-15 minutes.
4.
Remove the product rinse line from the drain.
5.
Using aseptic technique, hold the product line over a container, and allow
approximately 100 cc of product (RO) water to flow to the drain (or allow product
water to flow for 30 seconds).
6.
Maintaining aseptic technique, hold the product line over the sterile container
with the product water flowing.
7.
Fill the container with RO product water. (If using commercial samplers, follow
the manufacturer's instructions.)
8.
Aseptically cap the specimen container immediately.
9.
Label the specimen appropriately with:
•
Test to be performed - "culture / colony count"
•
Sample source - product water/RO, machine serial number and room location
•
Time and date sample obtained
•
Person who obtained specimen
•
Any other pertinent information or procedures your facility or lab requires
10.
The samples should be assayed within 30 minutes of obtaining or refrigerated at
5
°
C immediately and assayed within a 24-hour period.
Summary of Contents for MILLENIUM OSMONICS MILLENIUMTM REVERSE OSMOSIS UNIT
Page 2: ......
Page 6: ...14693e D D 31 March 2003 This page intentionally left blank ...
Page 24: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 1 18 General Information NOTES ...
Page 36: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 3 5 5 System Operation NOTES ...
Page 67: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 4 30 Routine Maintenance NOTES ...
Page 80: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 5 13 Troubleshooting NOTES ...
Page 98: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 6 17 Drawings NOTES ...
Page 105: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 7 6 Spare Parts List NOTES ...
Page 106: ......