Section 8-12 - Functional Checks after FRU replacement (Debrief Guideline)
8 - 241
8-12-92
Replacement of the U-WIFI Module Assy
8-12-93
Replacement of the HDU 23.8 inch Monitor LOGIQ Exx
8-12-94
Replacement of the OPIO HUB To Monitor Power Cable
8-12-95
Replacement of the OPIO HUB To Monitor USB 2.0 Cable
Functional Check
Debrief
Script
Equipment passed all required checks and is ready for
use.
4-3-8 "B-Mode functional checks"
4-3-10 "System CFM and PWD functional checks"
4-3-20 "Basic measurement functional checks"
Functional Check
Debrief
Script
Equipment passed all required checks and is ready for
use.
4-3-8 "B-Mode functional checks"
4-3-10 "System CFM and PWD functional checks"
10-6-6 "Earth Leakage Current Test"
Functional Check
Debrief
Script
Equipment passed all required checks and is ready for
use.
4-3-8 "B-Mode functional checks"
4-3-10 "System CFM and PWD functional checks"
10-6-6 "Earth Leakage Current Test"
Functional Check
Debrief Script
Leakage Current measured at (record the value)
and meets allowable limits.
Equipment passed all required checks and is ready
for use.