Basic Service Manual
8 - 240
Section 8-12 - Functional Checks after FRU replacement (Debrief Guideline)
8-12-89
Replacement of the UECG Assy
8-12-90
Replacement of the VNAV Option Cable Set-U
8-12-91
Replacement of the 3DG DriveBay 2+ EU and Power Adapter Cable
Functional Check
Debrief
Script
Equipment passed all required checks and is ready for
use.
4-3-8 "B-Mode functional checks"
4-3-10 "System CFM and PWD functional checks"
4-3-14 "PW/CW Doppler mode functional checks"
4-3-20 "Basic measurement functional checks"
4-3-22 "ECG functional checks"
Utility->Applications->Settings->ECG->Show ECG tools.
Connect cables. Press Display ECG tab on Touch Panel.
Verify ECG signal displayed on the monitor.
Plug in the ECG cables with nothing attached so they pick
up noise.
4-3-24 "Volume Navigation with inside probe func
Functional Check
Debrief
Script
Equipment passed all required checks and is ready for
use.
4-3-7 "Probe/connector functional checks"
4-3-8 "B-Mode functional checks"
4-3-10 "System CFM and PWD functional checks"
4-3-20 "Basic measurement functional checks"
4-3-24 "Volume Navigation with inside probe func
Functional Check
Debrief
Script
Equipment passed all required checks and is ready for
use.
4-3-7 "Probe/connector functional checks"
4-3-8 "B-Mode functional checks"
4-3-10 "System CFM and PWD functional checks"
4-3-20 "Basic measurement functional checks"
Summary of Contents for LOGIQ Totus
Page 2: ......
Page 4: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual ii...
Page 5: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual iii...
Page 6: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual iv...
Page 643: ......