Section 8-5 - Replacing Cables
8 - 49
8-5-7
Replacement of the Lower Keyboard To Touch Panel PWA Power & USB Cable
This table includes information specific to these instructions. For more information
for servicing the LOGIQ Totus" on page 8-5
8-2-7 "PPE Required During Service" on page 8-6
.
8-5-7-1
Preparations
1.) Power off/Shutdown the system as described in
1-7-2 "Lockout/tagout procedure" on page 1-25
2.) Remove the OPIO Top Cover Set-T as described in
8-7-2 "Replacement of the OPIO Top Cover
3.) Remove the UOPIO-T Assy as described in
8-7-10 "Replacement of the UOPIO-T Assy" on page
8-5-7-2
Removal Procedure
8-5-7-3
Installation Procedure
1.) Parts to be installed in reverse order of removal
2.) Perform “Functional Checks after FRU replacement (Debrief Guideline)” as described in
Table 8-25 Manpower, Time, Tools, and PPE
Manpower
Time
Tools
PPE
One person
Approximately 15 minutes
Phillips screwdriver
Cut resistant gloves
Knee Pad
Steps
Corresponding Graphic
1.
Disconnect the Lower Keyboard To Touch Panel
PWA Power & USB Cable from the LOWER OPIO
and the UPPER OPIO. Unscrew 1 screw to remove
the Lower Keyboard To Touch Panel PWA Power &
USB Cable.
3.
Remove the plastic clamp from the Lower Keyboard
To Touch Panel PWA Power & USB Cable.
Summary of Contents for LOGIQ Totus
Page 2: ......
Page 4: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual ii...
Page 5: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual iii...
Page 6: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual iv...
Page 643: ......