Safety Precautions
LOGIQ E9
–
User Guide
4-29
Direction 5454884-100 English
Rev. 1
Declaration of Immunity (continued)
Table 4-6:
Declaration of Immunity
Immunity Type
Equipment
Capability
Regulatory
Acceptable Level
EMC Environment
and Guidance
IEC 61000-4-2
Static discharge (ESD)
±
9 kV contact
±
12 kV air
±
9 kV contact
±
12 kV air
Floors should be wood,
concrete, or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should
be at least 30%.
Mains power quality
should be that of a typical
commercial and/or
hospital environment. If
the user requires
continued operation
during power mains
interruptions, it is
recommended that the
system be powered from
a UPS or a battery.
NOTE: UT is the a.c.
mains voltage prior to
application of the test
level.
Power frequency
magnetic fields should be
at levels characteristic of
a typical location in a
typical commercial and/or
hospital environment.
Separation distance to
radio communication
equipment must be
maintained according to
the method below.
Interference may occur in
the vicinity of equipment
marked with the symbol:
IEC 61000-4-4
Electrical fast
transient/burst
±
2 kV for mains
±
1 kV for SIP/SOP
±
2 kV for mains
±
1 kV for SIP/SOP
IEC 61000-4-5 Surge
Immunity
±
1 kV differential
±
2 kV common
±
1 kV differential
±
2 kV common
IEC 61000-4-11
Voltage dips, short
interruptions and
voltage variations on
mains supply
< 50
T
(> 95% dip) for
0.5 cycle;
400
T
(60 0ip) for 5
cycles;
700
T
(30 0ip) for 25
cycles;
< 50
T
(>95% dip) for 5
sec
< 50
T
(> 95% dip) for
0.5 cycle;
400
T
(60 0ip) for 5
cycles;
700
T
(30 0ip) for 25
cycles;
< 50
T
(>95% dip) for 5
sec
IEC 61000-4-8
Power frequency (50/60
Hz) magnetic field
6 A/m
6 A/m
IEC 61000-4-6
Conducted RF
3 V
RMS
150 kHz - 80 MHz
3 V
RMS
150 kHz - 80 MHz
IEC 61000-4-3
Radiated RF
3 V/m
80 MHz - 2.5 GHz
3 V/m
80 MHz - 2.5 GHz
IEC 61000-3-2
Harmonic Disturbance
Emissions
IEC Limits
230Vac
IEC 61000-3-3 Voltage
Fluctuations Emissions
IEC Limits
230Vac 50Hz, 100Vac
50Hz
NOTE: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects, and people. If noise generated from other electronic
equipment is near the probe’s center frequency, noise may appear on the image. Good power line isolation
is required.
Summary of Contents for LOGIQ E9
Page 8: ...i 6 LOGIQ E9 User Guide Direction 5454884 100 English Rev 1 ...
Page 12: ...i 10 LOGIQ E9 User Guide Direction 5454884 100 English Rev 1 ...
Page 126: ...Performing an Exam 2 40 LOGIQ E9 User Guide Direction 5454884 100 English Rev 1 ...
Page 202: ...After the Exam is Over 3 76 LOGIQ E9 User Guide Direction 5454884 100 English Rev 1 ...
Page 250: ...Safety 4 48 LOGIQ E9 User Guide Direction 5454884 100 English Rev 1 ...
Page 254: ...Index 4 LOGIQ E9 User Guide Direction 5454884 100 English Rev 1 ...
Page 255: ......
Page 256: ......