Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
8-4
Section 8-3 - Loading the System Software
Section 8-3
Loading the System Software
NOTE:
System Software may also be referred to as the Base Image or Ghost.
NOTE:
Before loading the system software, please ensure that the power can be continuously supplied
and there is no risk of power cut off during loading procedure.
There are two methods to load the system software:
•
Load the system software with DVD. Refer to section
8-3-1 "Loading Base Image Software with
from
to
•
Load the system software with USB memory stick. Refer to Refer to section
System Software with USB Memory Stick" on page 8-10
from
.
NOTE:
When upgrading the software with DVD or USB, please remove the system from the docking
cart. Otherwise, the software update might fail.
WARNING
WARNING
While the software install procedure is designed to preserved data, you should save
any patient data, images, system setups and customer presets to CD, DVD, USB Flash
Drive, or USB Hard Disk before doing a software upgrade.