Disinfecting probes (continued)
1. Prepare the germicide solution according to the
manufacturer's instructions. Be sure to follow all precautions
for storage, use and disposal.
2. Place the cleaned and dried probe in contact with the
germicide for the time specified by the germicide
manufacturer. High-level disinfection is recommended for
surface probes and is required for endocavitary and
intraoperative probes (follow the germicide manufacturer's
recommended time).
Probes for neuro surgical intra-operative use must NOT be
sterilized with liquid chemical sterilants because of the
possibility of neuro toxic residues remaining on the probe.
Neurological procedures must be done with the use of
legally marketed, sterile, pyrogen free probe sheaths.
3. After removing from the germicide, rinse the probe following
the germicide manufacturer's rinsing instructions. Flush all
visible germicide residue from the probe and allow to air dry.
CAUTION
CREUTZFIELD-JACOB DISEASE
Neurological use on patients with this disease must be
avoided. If a probe becomes contaminated, there is no
adequate disinfecting means.
Biological
Hazard
Summary of Contents for LOGIQ 7
Page 4: ...i 2 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 This page intentionally left blank ...
Page 16: ...i 14 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 28: ...Introduction 1 12 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 140: ...Preparing for an Exam 4 28 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 306: ...Scanning Display Functions 6 42 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 418: ...Abdomen and Small Parts 8 18 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 644: ...Cardiology 10 128 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 688: ...Urology 12 8 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 698: ...Pediatrics 13 10 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 882: ...Customizing Your System 16 106 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 928: ...Probes and Biopsy 17 46 LOGIQ 7 Online Help Direction 2392536 100 Rev 1 ...
Page 974: ...User Maintenance 18 46 LOGIQ 7 Online Help Direction 2392536 100 Rev 1 ...
Page 983: ...GE Medical Systems ...
Page 984: ...1 2 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...