8-160
Section 8-6 - Mechanical Parts- Replacement Procedures
D
IRECTION
5771498-100, R
EVISION
6
V
ENUE
™ S
ERVICE
M
ANUAL
P R E L I M I N A R Y
8-7-12
MPB To Peripherals Harness - FRU Cable Replacement Procedure
8-7-12-1
Tools
•
Appropriate Phillips screwdriver
•
Flat-head screwdriver
8-7-12-2
Time Required
15 min
8-7-12-3
Preparations
1.) Shut down the Venue™ ultrasound unit, as described in
.
2.) Make sure the On/Off power switch is set to Off.
3.) Make sure the system is standing securely on a level surface, with the wheels in the locked position.
8-7-12-4
MPB To Peripherals Harness - FRU Cable Removal Procedure
1.) Perform MPB removal procedure.
2.) Release the cable from the plastic cable clips using a screwdriver
•
Figure 8-166 Disconnecting Cable Clips