GE H
EALTHCARE
D
IRECTION
5394141, R
EVISION
5
LOGIQ™ P5 S
EVICE
M
ANUAL
Section 10-3 - Maintenance Task Schedule
10-3
Table 10-2 Customer Care Schedule
Service at Indicated Time
Daily
Monthly Quarterly
Per Facilities
QA Program
Notes
Clean Probes
•*
* or before each use
Clean Probe Holders
•
Clean Air Filter
•
more frequently depending on
your environment
Inspect AC Mains Cable
•
Mobile Unit Check Weekly
Inspect Cables and Connectors
•
Clean Console
•
Clean Monitor and Touch Panel
•
Inspect Wheels, Casters, brakes and
Swivel Locks
•
Mobile Unit Check Daily
Check Control Panel Movement
•
Mobile Unit Check Daily
Console Leakage Current Checks
•
also after corrective
maintenance
Peripheral Leakage Current Checks
•
also after corrective
maintenance
Surface Probe Leakage Current Checks
•
also after corrective
maintenance
Endocavity Probe Leakage
Current Checks
•
also after corrective
maintenance
Transesphongeal Probe Leakage
Current Checks
•
also after corrective
maintenance
Surgical Probe Leakage
Current Checks
•
also after corrective
maintenance
Measurement Accuracy Checks
•
also after corrective
maintenance
Probe/Phantom Checks
•
also after corrective
maintenance
Functional Checks
•
also after corrective
maintenance