Chapter 7: Repair Procedures
Right Rail Assembly Repair Procedures
Service Manual
7-23
7.4.3 Right Rail Internal Repairs
7.4.3.1 Removing the Inner Rail Assembly
“Figure 7-24 Removing the Lift Motor”
,
“Figure 7-25 Right Rail Disassembly”
, and
.
1.
Using a 2.5 mm hex key, remove the 2 screws located at either end of the upright in the rear dovetails
that secure belt channel to the inside of the upright. Using a 3 mm hex key, remove the side roller
bracket. Remove the lift motor coupler components if still attached to the rail assembly. Pull the entire
rail assembly up out of the upright.
2.
Remove the pinch guard by sliding it up then pulling it off.
3.
Remove the nut and washers at the bottom of the belt channel that secure the heater door opening rod
to the bracket and remove the assembly.
FIGURE 7-25. Right Rail Disassembly
Lift rail
Belt channel
Motor bracket
Right upright
Belt block
Rollers
Buoyancy springs
Access hole
Side roller
bracket
Pinch guard
Summary of Contents for Giraffe OmniBed Carestation CS1
Page 4: ...RH 2 Service Manual ...
Page 24: ...Service Language Disclaimer viii Service Manual ...
Page 40: ...Safety and Regulatory Information User Responsibility xxiv Service Manual ...
Page 44: ...Chapter 1 Functional Description Control Board 1 2 Service Manual FIGURE 1 1 Block Diagram ...
Page 74: ...Chapter 3 Calibration and Maintenance Line Voltage Calibration 3 12 Service Manual ...
Page 82: ...Chapter 4 Understanding the Service Interface Upgrade 4 8 Service Manual ...
Page 110: ...Chapter 5 Troubleshooting Diagnostic Readings 5 28 Service Manual ...
Page 211: ...Chapter 7 Repair Procedures Display Harness Connections Service Manual 7 75 ...
Page 212: ...Chapter 7 Repair Procedures Display Harness Connections 7 76 Service Manual ...
Page 294: ...Chapter 8 Illustrated Parts Wiring Diagrams 8 82 Service Manual ...
Page 296: ...Appendix A Compartment and Skin Probe Characteristics A 2 Service Manual ...
Page 304: ...Appendix C RS 232 Serial Data Nurse Call C 4 Service Manual ...