GE
D
IRECTION
FQ091019, R
EVISION
2
V
IVID Q
N S
ERVICE
M
ANUAL
8-22
Section 8-2 - External Component Replacement Procedures
8-2-6
LCD Display Cover Hinges Replacement Procedure
NOTE:
This section describes the replacement procedure for the LCD Display Cover hinges
only
.
Refer to
. (Part No. KTZ280195; one pair of hinges).
8-2-6-1
Tools
Phillips screwdriver; Loctite 242 glue
8-2-6-2
Time Required
15 minutes
8-2-6-3
Preparations
Shut down the Vivid q N ultrasound unit, as described in
8-2-6-4
LCD Display Cover Hinges Removal Procedure
1) Remove the bearing handle, as described in the
"Bearing Handle Removal Procedure" on page 8-2
.
2) Remove the Control Panel and Keyboard as described in the
Removal Procedure" on page 8-10
3.) Slide to open the two connector latches and disconnect the LCD Flex ribbon cable from BEP -
.
4.) Unscrew and remove the four screws (two on either side - as shown in
) that secure the
Rear Cover and Latch Assembly to the Bottom Assembly.
Figure 8-25 Connector Latches
Figure 8-26 Rear Cover and Latch Assembly Securing Screws