Technical Specifications 8
8 Technical
Specifications
Regulation
Maximum voltage, current and power with automatic cross-
over at preset limits
Output range
Voltage:
0–600 V DC
Current:
0–400 mA
Power:
0–100
W
Programming range
Voltage:
6–600 V DC
Current:
1–400
mA
Power:
1–100 W
Time:
00:01–500 h
Volthour:
1–500 000 Vh
Output resolution
Voltage:
1 V
Current:
100
μA,
0–39.9
mA
1
mA,
40–400
mA
Power:
1W
Programming resolution
Voltage:
1 V
Current:
1 mA
Power:
1 W
Time:
1
min,
00:01–99:59,
1h, 100–500 h
Volthour:
1 Vh, 1–9999 Vh,
100 Vh, 10.0–99.9 kVh
1 kVh, 100–500 kV
Accuracy
Voltage:
1%, ±3 V
Current:
1%, ±1 mA
Power:
2%,
±1
W
Timer:
0.1% ±1 min, 00:01–99:59 h
0.1%, ±10 min, 100–500 h
Line regulation
< 0.2%
Load regulation
< 1% at load change 10–90% of maximum load
Ripple
< 1% at 600 V
Short term stability
< 0.2% /10 h after warm up
Long term stability
< 1% /year
Start current check
Resistance not greater than 0.2 Mat 40 V
(current less than 200 μA). Can be disabled
Sudden load change check
Ground leakage check
Leakage not greater than 500 μA
Output protection
Fully protected against any overload conditions
Recovery after power failure
Duration < 8 s, The program continues automatically
Duration ≥ 8 s, The program continues after manual restart
27