16
Electrophoresis Power Supply EPS 301 User Manual 18-1130-18 Edition AC
ENGLISH
FRANCE
DEUT
CH
ESP
AÑOL
IT
ALIANO
8 Technical specifications
Regulation
Maximum voltage and current with automatic cross-over
at preset limits
Output mode
Contante Voltage: 0–300 V DC,
Contante
Current:
0–400
mA
Maximum Power Output
80 W
Programming range
Voltage: 5–300 V DC
Current:
10–400
mA
Time: 00.01–23.59 h or off (- - - -)
Output resolution
Voltage: 1 V
Current:
1
mA
Programming
resolution
Voltage:
1
V
Current:
1
mA
Time:
1
min
Accuracy
Voltage: 4%, ± 2 V
Current: 4%, ± 4 mA
Timer: 0.1% ± 1 min
Line regulation
< 0.5%
Load regulation
< 1% at load change 10–90% of maximum load
Ripple
< 3% at 300 V
Short term stability
< 0.2% /10 h after warm up
Long term stability
< 1% /year
Output protection
Fully protected against any overload conditions
Recovery after power failure
The program continues automatically
Ambient operating temperature
4–40 °C
Ambient operating humidity
0–95%
Ambient operating pressure
68–106 kPa, maximum altitude of 2000 m
Mains requirements
100–120 V/220–240 V; 50/60 Hz
Power consumption
Max 120 W
Dimensions (WxDxH)
250 x 215 x 95 mm
Weight 3.0
kg
8 Technical
specifications