User Manual
Druck DPI620 Genii
Procedures (CH1/CH2): Current (source)
8.3
Do the procedure as follows:
1.
Connect the applicable calibration equipment
(Ref: Table 8-1):
•
CH1/CH2 (24 mA range): (Ref: Section 2.4 and Section
2.9).
•
CH2 (-24 mA range only): (Ref: Section 2.4 and Section
2.9).
2.
Let the equipment get to a stable temperature
(minimum: 5 minutes since the last power on).
3.
Use the calibration menu (Ref: Section 8.1) to do a two-
point calibration (0.2 mA and FS):
•
CH1 (one range): 24 mA.
•
CH2 (two ranges): 24 mA (reverse) and 24 mA
(forward).
Note
: Input positive values for the forward and reverse
calibration.
06 June 2013
K0541 issue 1
Page 144 of 175