CL 5
Issue 01, 04/05
2-9
2.3 Working with the
Configuration Display
The Configuration display is accessed by pressing
.
With this screen displayed, you can adjust most CL 5
controls.
VIEW
MODE
In base-model units this parameter
offers 4 settings. In A-Scan equipped instruments this
parameter allows the measurement and live A-Scan to
be displayed. See Section 4.1 to select the view to be
displayed.
DATA RECORDER
When the optional data recorder
is installed, setting this parameter to ON activates the
file recording capability and the data recorder file
display.
MINIMUM ALARM
Enables and sets the minimum
alarm value. (Section 3.4)
MAXIMUM ALARM
Enables and sets the maximum
alarm value. (Section 3.4)
NOMINAL THICKNESS
Set the nominal thickness
value that is then used to calculate and the differential
and rate of reduction measurement values when
selected by the VIEW MODE parameter.
(Section 3.2.3)
RESOLUTION
Controls the format (number of
decimal places) of the thickness measurement
displayed, printed in a report, and stored in a Data
Recorder file. (Section 3.2)
UPDATE RATE
Rate at which the measurement
reading is updated. (Section 3.2.2)
BACKLIGHT
Turns the displays backlight feature to
ON, OFF, or AUTO (AUTO switches backlighting on
whenever a key is pressed or the probe is coupled)
CONTRAST
Sets the displays contrast level (select
then press or to change contrast)
LANGUAGE
Sets the instrument displays language
(Section 3.2)
UNITS
Sets the unit of measurement to inches or
millimeters (Section 3.2)
Understanding the Keypad, Menu System, and Displays
Working with the Configuration Display Module
CL 5
Issue 01, 04/05
2-9
2.3 Working with the
Configuration Display
The Configuration display is accessed by pressing
.
With this screen displayed, you can adjust most CL 5
controls.
VIEW
MODE
In base-model units this parameter
offers 4 settings. In A-Scan equipped instruments this
parameter allows the measurement and live A-Scan to
be displayed. See Section 4.1 to select the view to be
displayed.
DATA RECORDER
When the optional data recorder
is installed, setting this parameter to ON activates the
file recording capability and the data recorder file
display.
MINIMUM ALARM
Enables and sets the minimum
alarm value. (Section 3.4)
MAXIMUM ALARM
Enables and sets the maximum
alarm value. (Section 3.4)
NOMINAL THICKNESS
Set the nominal thickness
value that is then used to calculate and the differential
and rate of reduction measurement values when
selected by the VIEW MODE parameter.
(Section 3.2.3)
RESOLUTION
Controls the format (number of
decimal places) of the thickness measurement
displayed, printed in a report, and stored in a Data
Recorder file. (Section 3.2)
UPDATE RATE
Rate at which the measurement
reading is updated. (Section 3.2.2)
BACKLIGHT
Turns the displays backlight feature to
ON, OFF, or AUTO (AUTO switches backlighting on
whenever a key is pressed or the probe is coupled)
CONTRAST
Sets the displays contrast level (select
then press or to change contrast)
LANGUAGE
Sets the instrument displays language
(Section 3.2)
UNITS
Sets the unit of measurement to inches or
millimeters (Section 3.2)
Understanding the Keypad, Menu System, and Displays
Working with the Configuration Display Module
Summary of Contents for CL 5
Page 11: ...CL 5 Issue 01 04 05 1 1 General Information 1 CL 5 Issue 01 04 05 1 1 General Information 1 ...
Page 29: ...CL 5 Issue 01 04 05 3 1 Setting Up the CL 5 3 CL 5 Issue 01 04 05 3 1 Setting Up the CL 5 3 ...
Page 45: ...CL 5 Issue 01 04 05 4 1 Measuring Thickness 4 CL 5 Issue 01 04 05 4 1 Measuring Thickness 4 ...
Page 63: ...CL 5 Issue 01 04 05 6 1 I O Features 6 CL 5 Issue 01 04 05 6 1 I O Features 6 ...
Page 73: ...CL 5 Issue 01 04 05 7 1 Specifications 7 CL 5 Issue 01 04 05 7 1 Specifications 7 ...
Page 81: ...CL 5 Issue 01 04 05 8 1 Maintenance 8 CL 5 Issue 01 04 05 8 1 Maintenance 8 ...
Page 83: ...CL 5 Issue 01 04 05 9 1 Appendix 9 CL 5 Issue 01 04 05 9 1 Appendix 9 ...
Page 89: ...CL 5 Issue 01 04 05 10 1 Index 10 CL 5 Issue 01 04 05 10 1 Index 10 ...